ABF Section Dialog

This page describes the contents of the Axon Instruments ABF section dialog.

Controls

Enable

If checked, the section is enabled. If unchecked, the section is not used.

Start

The start time of the section in seconds. The corresponding sample number is displayed below the start time.

End

The end time of the section in seconds. The corresponding sample number is displayed below the end time. The duration of the selected data is displayed to the right of the end time and the corresponding sample number is displayed below the duration.

Buttons

OK
Accept the current settings and close the dialog.
Cancel
Reject any changes to the settings and close the dialog.
Help
Bring up this text.

Notes

Time and Split Clock

If your ABF file is recorded with two sampling intervals, the dialog operates as if the data were acquired only at the first sampling rate. For example, suppose your episode has a total of 200 points consisting of 100 points acquired at an interval of 0.1ms/sample, followed by 100 points acquired at 1ms/sample. The episode represents 10ms of recording at 0.1ms/sample followed by 100ms of recording at 1.0ms/sample. If you specify a start time of 0s and an end time of 0.015s, the 15ms time interval is interpreted at the first sampling rate of 0.1ms/sample, resulting in 150 samples. These 150 samples actually correspond to 60ms, that is, 100 samples at 0.1ms/sample, followed by 50 samples at 1ms/sample.